Minor equipment repair manual



Minor equipment repair manual

Instrument sterilizer
Change kettle element
Remove calcuim deposit with tezaab
Change plug ceramic connector wire
Washer with kettle element

Bp apparatus
Change bp bulb and bladder
Bp bulb with valve
Change broken glass tube
Fill mercury if lost mercury

Horizontal and vertical autoclave
Change geyser element

Microscope
Clean lens with spirit eyepiece and objective
Change any broken objective
I suggested may weld microscope to avoid collision
Repair microscope lamp 

Colorimeter
Caliberate blank
Change fused lamp must be 6v 10watt halogen same for analyser also

Semi auto analyser
Change tubing
Change fused lamp
Order test kits
Order pm kits
Wash flowcell after each test by aspirating distilled water
May remove flowcell
Keep test tube in flowcell container
Fit computer ribbon connector for filter wheel

Xray machine
Change xray machine control panel fuses
Remove two main switch fuses make single phase
Remove grid Disengage bucky motor
Make wooden table

Suction machine
Change plug
Repair wires
Empty the bottles
Get motor serviced
Repair tubing by inserting metal wire

Ultrasound
Keep one probe on front other on back
Dip probe in oil and switch on

Anesthesia machine
Close air hole with sheetmetal and gi clamp or using surgical tape.
Remove oxygen and nitrous metal neck
Fit rubber hose and lpg regulator 

Ot light
Pour battery water in lead acid battery inverter
Do external wiring if wire blown
Change fused bulbs and halogen lamps

Radiant heat warmer change room heater connecting metal pieces

Ecg machine
Check continuity patient cable and solder
Any wire broken
Change on off switch lead acid battery keep under charge
Defibrillator also
Clean stylus with spirit and cotton 

Hot air oven and lab incubator short broken heater coil elements inside walls of cabinet 

Electrosurgery order metallic patient plate
Repair broken patient cable solder phonoplug used with ahuja amplifier

Fridge and air conditioner and water cooler
Pour oil on small cylinder tube
Seal with surgical tape broken compressor port.

Needle destroyer
Align the blades
Scratch and clean with emery paper the blades.

Thermal printer fax clean print head with spirit.
Centrifuge bypass safety switch change carbon brush with mixie carbon brush and change fan regulator 
Dental chair unit
Send handpieces
Change water bottle
Change halogen lamp
Oil the hand piece with coconut oil
Rewind air compressor or order new cycle shop compressor
Treadmill test
Oil the treadmill
Pulse oximeter
Order new probe
Stick old broken probe with superglue
Ventilator
Some air hole closed max ventilator mohali
Oil cieling fan

Mahesh n